The Generic LIMS CSV Interface (Q12351) can be used to import results from the Hach TU5200 Benchtop Laser Turbidimeter CSV Data log.

Configure your TU5200 to log data to CSV:

  1. At the main reading screen, push Options>Reading Setup>Data Log Setup>Send Data Format
  2. Set to CSV
  3. Make sure that Auto Store is on (On is the default).
  4. Setup the TU5200 to log data to NetDrive (Folder where the Data Log CSV files will be placed).  The instructions for setting the instrument up to a LAN connection are on page 16 of the TU5200 User Manual.

Configure the interface:

  1. Go to Configuration>Source Configuration

  1. Set the Import Folder to the location of your TU5200 CSV files.  The Data Log files are typically named :  DL_TU5200_sssssss_yyyy-mmm-dd_hh_mm_ss.csv where ssssss is the Instrument serial number (Ex DL_TU5200_1664014_2016-Oct-28_13_18_21.csv)
  2. Set the number of header rows to 5. 
  3. Map the Columns as follows:

NOTE: Analyte is set to Units as the file only contains Turbidity results.  This way the variable can be cross referenced as follows:

Related Articles
No Related Articles Available.

Article Attachments
No Attachments Available.

Related External Links
No Related Links Available.
Help us improve this article...
What did you think of this article?

poor 
1
2
3
4
5
6
7
8
9
10

 excellent
Tell us why you rated the content this way. (optional)
 
Approved Comments...
No user comments available for this article.
Created on 3/24/2017 12:23 PM.
Last Modified on 3/24/2017 2:52 PM.
Last Modified by Scott Dorner.
Article has been viewed 3444 times.
Rated 7 out of 10 based on 6 votes.
Print Article
Email Article